E-DWT Electronic Deadweight Tester Kits |
|
型 號:6531 and 6532 |
品 牌:美國FLUKE |
技術指標:To RPM4-E-DWT:12 V dc 1.2 A |
|
廊坊E-DWT Electronic Deadweight Tester Kits-6531 and 6532-美國FLUKE
A powerful, complete hydraulic pressure calibration system to cover a wide
workload
Fluke Calibration
improved the hydraulic pressure calibration process with the introduction of
E-DWT-H electronic deadweight tester. E-DWT-H is an electronic calibrator
designed to deliver traditional hydraulic deadweight tester performance with
digital measurement features and convenience. Some improvements offered by
E-DWT-H include:
·
No weights to load and unload, transport
or send out for calibration
·
Provides real-time pressure indication
with no need to know and correct for local gravity or ambient temperature
·
No piston-cylinder changes
·
No significant sensitivity to level or
vibration
·
Able to set and read any pressure value
exactly, no minimum increment limited by smallest available masses; perfect for
applications that require setting a nominal pressure precisely on the device
under test and measuring it, such as analog gauge calibration
·
Operates in any unit of measure, switching
easily from one unit to the next
The E-DWT-H
is at home in metrology and calibration labs, on the production floor or in the
field. It can operate with a wide selection of test mediums, including Sebacate
calibration fluid, mineral oil, water and other liquids. ± 0.02 % of reading
total one-year measurement uncertainty rivals the best laboratory deadweight
testers.
The E-DWT-H is
used in the lab or instrument shop or can be taken into the field for
performing in-situ calibrations and tests. This hydraulic pressure calibration
system combines a high performance electronic reference pressure monitor and
manually operated pressure generation and control hardware in a single, compact
and rugged package. E-DWT-H offers the ease of use and precision of continuous,
real time electronic pressure measurement with the simple and direct pressure
control of high quality manual pressure hardware. E-DWT is named “electronic
deadweight tester” because it is intended as a lightweight and easy to use,
modern alternative to conventional piston-cylinder and weight based deadweight
testers.
The E-DWT’s
electronic reference pressure monitor is a special version of RPM4, designated
RPM4-E-DWT. RPM4-E-DWT can be configured with one or two high precision quartz
reference pressure transducers (Q-RPTs) with ranges from 7 MPa (1 000 psi) to
200 MPa (30 000 psi).
The RPM4-E-DWT is
controlled locally by the operator using its front panel display, keypad and
optional foot pedal or remotely by a computer using ASCII character command
strings over its RS-232 interface.
RPM4-E-DWT uses an
AutoRange feature to automatically select the most appropriate Q-RPT and to
optimize the E-DWT-H setup to cover the desired range of operation. When two
Q-RPTs are included, a half-turn valve isolates and protects the Lo Q-RPT from
high pressure when the Hi Q-RPT is in use. Visual and audible indicators assist
the operator in setting the shut off valve correctly.
The E-DWT-H
includes the hardware necessary to fill and prime the system under test and
generate and precisely adjust pressure up to 200 MPa (30 000 psi).
The E-DWT-H can be
powered with an optional battery/charger pack for use in locations where AC
power is not readily available.
High performance calibration made simple
Fluke Calibration
electronic deadweight tester kits feature E-DWT-H configurations combined with
the accessories needed for a complete calibration system. Model 6531 options
feature E-DWT-H configurations with a single quartz reference pressure
transducer (Q-RPT) to offer percent of reading performance from 10 % to 100 %
of device full scale. Model 6532 options feature an additional Q-RPT to
maximize workload coverage, with percent of reading performance from 1 % to 100
% of device full scale. An included vacuum fill kit allows the E-DWT to be
filled with your choice of compatible fluid, while eliminating problematic air
pockets from the test circuit. Also included is a test station adapter to mount
just about any pressure device to be tested without using PTFE tape or
wrenches. 6531 and 6532 kits provide all you need to perform high level
hydraulic pressure calibration right out of the box.
·
Pressure ranges available to 200 MPa
(30,000 psi)
·
± 0.02 % of reading total one-year
accuracy, 0.025 % for two years
·
Low torque variable volume allows for
pressure generation and control up to 200 MPa (30,000 psi) with minimal
physical effort
·
Built-in priming system to fill system
with test fluid and purge unwanted air to assure smooth operation
·
Unique test port design eliminates the
need for PTFE tape or wrenches. Adapters included for 1/8 in, 1/4 in, 3/8 in
and 1/2 in NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x
1.5
·
Compatible with a wide range of liquids,
including water. Vacuum fill kit included to fill the E-DWT with any compatible
test medium
·
AutoTest™, AutoRange, ready/not ready
indicator, onboard data capture and storage, overpressure alarm, RS-232 remote
interface and other advanced features to simplify testing, improve safety and
prevent calibration errors and damage to equipment
·
Optional foot switch accessory allows
hands-free data collection when running AutoTests
·
Rechargeable battery pack and transport
case options for field operation
General
|
Power
requirements
|
To RPM4-E-DWT:
|
12 V dc 1.2 A
|
To AC to DC power supply:
|
100 V ac to 240 V ac, 50/60 Hz
|
Temperature
range
|
Storage:
|
- 20 °C to 70 °C
|
Operating:
|
10 °C to 40 °C
|
Relative
Humidity
|
Storage:
|
0 % to 100 %
|
Operating:
|
0 % to 70 %
|
Weight
|
1 Q-RPT:
|
12 kg (26 lb) approximate
|
2 Q-RPT:
|
14 kg (30 lb) approximate
|
Dimensions
|
E-DWT footprint:
|
41.4 cm W x 37.1 cm D (16.3 in x 14.6
in)
|
E-DWT height:
|
26.9 cm (10.6 in), 33.6 cm (13.2 in) to
max variable volume handle height
|
Pressure range
|
Up to 200 MPa (30k psi), depending on
configuration
|
Operating medium
|
All 6531 and 6532 E-DWT Kits are shipped
dry, standard preparation. Can be filled with di-ethyl-hexyl sebacate,
silicon oils, propylene glycol, fully fluorinated liquids, partially
fluorinated liquids, isopropyl alcohol and distilled water or mineral oil.
|
Reservoir
capacity
|
300 cc (18 in3)
|
|
|
Variable volume
displacement
|
3 cc (0.18 in3), 200 MPa (30,000 psi)
maximum
|
Filling and
priming pump displacement
|
3.7 cc (0.23 in3)
|
TEST pressure
connection
|
DH500 female, 200 MPa (30k psi) maximum
working pressure. Adapters included for 1/8 in, 1/4 in, 3/8 in and 1/2 in
NPT, 1/8 in, 1/4 in, 3/8 in and 1/2 in BSP, M20 x 1.5 and M14 x 1.5. Adapter
maximum working pressure is 140 MPa (20k psi). Note: DH500 is a gland and
collar type fitting for 6 mm (1/4 in) coned and left hand threaded tubes
equivalent to AE F250C, HIP HF4, 9/16-18 UNF, etc.
|
Pressure Limits
|
Maximum Working
Pressure:
|
Range of RPM4-E-DWT monitor’s Hi Q-RPT
|
Maximum priming
pump pressure:
|
700 kPa (100 psi)
|
Maximum Working
Pressure:
|
with Lo Q-RPT selected: Range of
RPM4-E-DWT monitor’s Lo Q-RPT
|
Communication
ports:
|
RS-232 (COM1, COM2)
|
·
Pressure
measurement
|
Warm up time
|
15 minute temperature stabilization
recommended from cold power up
|
Normal Operating
Temperature Range
|
10 °C to 40 °C (50 °F to 104 °F)
|
Resolution
|
Default: 0.01 % of active range
User adjustable to 1 ppm of Q-RPT maximum or 10 ppm of active AutoRange,
whichever is larger
|
Precision1
|
± 0.018 % of reading or 0.0018 % of
Q-RPT span, whichever is greater
|
Predicted
Stability2
|
|
One year:
|
± 0.0075 % of reading
|
Two year:
|
± 0.015 % of reading
|
Measurement
Uncertainty3
|
|
One year:
|
± 0.02 % of reading or 0.002 % of Q-RPT
span, whichever is greater
|
Two year:
|
± 0.025 % of reading or 0.0025 % of
Q-RPT span, whichever is greater
|
·
1. Combined linearity, hysteresis, and
repeatability. Precision does not include stability or calibration reference
uncertainty.
2. Predicted Q-RPT measurement stability limit (k=2) assuming regular use of
AutoZero function and short term stability between rezeroing.
3. Maximum deviation of the Q-RPT indication from the true value of applied
pressure including precision, predicted stability with rezeroing, temperature
effect from 10 oC to 40 oC and calibration uncertainty (assumes calibration
reference uncertainty of ±0.005 % of reading, k=2), combined and expanded (k=2)
following the ISO “Guide to the Expression of Uncertainty in Measurement.
型號
|
描述
|
6531-7M
|
E-DWT-H A7M, 0.7 to 7 MPa, 100
to 1,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6531-14M
|
E-DWT-H A14M, 1.4 to 14 MPa,
200 to 2,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6531-20M
|
E-DWT-H A20M, 2 to 20 MPa, 300
to 3,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6531-40M
|
E-DWT-H A40M, 4 to 40 MPa, 600
to 6,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6531-70M
|
E-DWT-H A70M, 7 to 70 MPa,
1,000 to 10,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6531-140M
|
E-DWT-H A140M, 14 to 140 MPa,
2,000 to 20,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6531-200M
|
E-DWT-H A200M, 20 to 200 MPa,
3,000 to 30,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6532-70M
|
E-DWT-H A70M/A7M, 0.7 to 70
MPa, 100 to 10,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6532-140M
|
E-DWT-H A140M/A14M, 1.4 to 140
MPa, 200 to 20,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
6532-200M
|
E-DWT-H A200M/A20M, 2 to 200
MPa, 300 to 30,000 psi
Includes:
· E-DWT-H
Electronic Deadweight Tester with power supply and operating manual
· Vacuum
liquid fill kit
· Test
station quick-mount fitting
· NPT
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· BSP
adapters: 1/8 in, 1/4 in, 3/8 in and 1/2 in
· Metric
adapters: M20 x 1.5 and M14 x 1.5
|
通用配件:
配件
|
描述
|
Case
|
加固且可重復使用的模制裝運外殼(部件號3338097)
|
電池組/充電器
|
帶有充電器的12 VDC電池(部件號3071049)
|
腳踏開關
|
遠程[ENTER]腳踏開關(部件號3070940)
|
COMPASS
for Pressure
|
自動執(zhí)行測試、數據采集、報告與設備資產管理的軟件
|
,,
|